Print this page.

GER 2600

Spectral Range350 nm to 2500 nm
Channels640
Linear Arrays(1) 512 Si
(1) 128 PbS
Bandwidth   1.5 nm:
                  11.5 nm:
350 nm to 1050 nm
1050nm to 2500 nm
Scan Time50 ms and up (selectable)
FOV3° standard, 10° option
23° option with fiber optic
Head Size320 mm x 335 mm x 114 mm
11.25" x 12" x 4.5"
Weight5.3 kg, 11.5 lbs.
Battery Type(1) 12 VDC
Battery Life4 hours
Digitization16 bit
Wavelength Repeatability+0.1 nm
Noise Equivalent Radiance0.5 s integration time
   400 nm: 6.2 x 10-9W · cm-2 ·nm-1· sr-1
   700 nm: 3.1 x 10-9W · cm-2 ·nm-1· sr-1
   900 nm: 9.3 x 10-9 W · cm-2 ·nm-1 · sr-1
   1500 nm: 9.0 x 10-9 W · cm-2 ·nm-1 · sr-1
   2200 nm: 9.0 x 10-9 W · cm-2 ·nm-1 · sr-1
Maximum Radiance Levels13 ms integration time
700 nm: 1.5 x 10-4W · cm-2 ·nm-1· sr-1
Radiometric Calibration Accurancy
(Traceable to NIST)
      400nm : ± 5%
      700nm : ± 4%
    1000nm : ± 5%
    2200nm : ± 7%
Dark Current Correction
UV-NIR-SWIR
automatic
Spectrum Averagingselectable
Humidity10 to 90% RH (non-condensing)
Temperature-10° to 50°C
SightingLaser

The GER 2600 is supplied in a sturdy, weather-tight field case, with batteries, battery charger and cables. System components can also be easily carried in a lightweight backpack.
  The GER 2600 system offers the user the greatest possible ease of operation for hand-held, back pack or tripod operation. Simple menu-driven programs control the set-up, acquisition, and data manipulation functions, allowing the user to be effective within an hour of introduction to the instrument.  Generated data are displayed on screen in real-time for immediate interpretation and evaluation.  Data is stored in ASCII format for easy transfer to other software programs. 
   Additionally, the GER 2600 spectroradiometer can be operated with the sub-notebook computer which provides real-time display and hard disk transfer.  By using the two different modes of operation, the operator can confirm the data integrity while maximizing collection efficiency.
Our technical team is ready to assist in evaluating your requirements to determine the best instrument for your specific application, whether it be in a remote sensing, agricultural, or industrial related field.

GER Corporation 16 Bennett Common Millbrook, NY 12545 USA
Phone: (845)677-6100 Fax: (845)677-6106 Email: info@ger.com Web: www.ger.com
©Copyright GER 1998-2000 all rights reserved.